Home > News > Company News

Test method for transfusion nursing package

2015-10-19 14:29:15 Zibo Chuangqi medical supplies Co. Ltd. Read
   We know that the product in the transfusion nursing package is needed to be tested, and only qualified products can be placed in the transfusion nursing package, which requires us to understand the test method:
   One, infusion care package in the medical pad
   Appearance:
   1,normal eyesight observation, should be consistent with the surface should be clean, smooth, cutting edge tidy, no broken edge hole, oil stains.
   2, size: with an accuracy of + 1mm scale measurement, should be in accordance with the deviation should be no more than 5%.
   3, impermeability: according to the method YY/T050 of test, the results should be consistent with shall comply with the YY/T0506.2 surgery single standard performance products non critical region, is more than or equal to 10cmH20.
   4, tensile strength, dry state: test according to YY/T0506.3 method, the results should be consistent with a dry state, shall comply with the YY/T0506.2 surgery single standard performance products non critical region, is more than or equal to 15N.
   Two, the nursing package transfusion tourniquet
   Appearance: normal eyesight observation, should be consistent with the surface should be clean, no obvious wrinkles and trimming neat, no pollution, uniform thickness.
   2, size: with an accuracy of + 1mm scale measurement, should be in accordance with the deviation should be no more than 3%.
   Three, infusion paste
   1, appearance: To observe normal eyesight, should be consistent with the surface should be flat, no pollution, no damage. Shall not be separated from the phenomenon of partial departure.
   2, size: with an accuracy of + 1mm scale measurement, should be in accordance with the length deviation should be no more than 5%, the width of the deviation should be no more than 2%.
   3, the peel strength test method: according to YY / T0148 in time, should be consistent with not less than 1.0N/cm.
   4, the adhesion test method of YY / T0148: according to the time, should meet should be no more than 2.5mm.
   Four, medical cotton swab
   1, size: allow deviation + 10mm.
   2, material: cotton swab used for medical cotton should be in accordance with the requirements of YY0330.
   3, appearance: should be clean, no stain, stain and foreign body, no smell, tasteless.
   Five, medical iodine cotton swab
   1, appearance: To observe normal eyesight, should meet should be clean, no stains, stains and foreign body, odorless, tasteless.
   2, size: with an accuracy of + 1mm scale measurement, should be in accordance with the allowable deviation of 10mm.
   3, material: according to the provisions of the YY0330 test, should be in accordance with the cotton swab used for medical cotton wool should be in accordance with the requirements of YY0330.
   In the infusion care packages for testing, for these knowledge we should understand, and only in the future use of us can be more assured.